VENTILATOR PARAMETERS |
RANGE |
Tidal Volume (at C mode) |
50~1200 ml |
Flow (at A/C, SIMV mode) |
Low, Med, High |
Breath Rate |
5-60 bpm, 1~12 bpm for SIMV |
%02 |
45%~90% |
Spont Flow |
0~5 L/Min |
Sigh |
1 per 0~100 breath cycle |
Inspiratory time |
0.5~3 s |
Trigger Level |
-10~0 hpa |
PIP |
0~60 hpa |
Silence |
60s |
High Pressure Alarm Limit |
20~60hpa (to be set) |
Low Pressure Alarm Limit |
2~20hpa (to be set) |
Alarm Indicator |
Flashing Red Indicator |
Over Pressure Relief |
<60 hpa |
Main Circuit Pressure (Paw) |
-2~10kPa |
VENTILATOR PARAMETERS |
RANGE |
Tidal Volume |
50~2000 ml |
Breath Rate |
1~100 bpm |
% 02 |
45%~100% |
Sigh |
1~10 per 100 breath cycle |
Trigger Level |
-20~0 hpa |
PIP |
0~60 hpa |
Silence |
60s |
Data Display |
Tidal Volume, 02%, Sigh, Respiratory Rate, Flow, Pressure (Synchronized Trigger, Peak Airway, CPAP) |
Curve Display |
Airway Pressure Time Waveform Display |
Alarms |
High/Low Airway Pressure Alarm, Lack of Gas/Power Alarm, Parameter Error |
Other Display Supply |
Lock, Charge Indicator, Inspiratory/Expiratory Cycle, DC/AC Power Supply |